9. MAINTENANCE

9a. Replacing the Batteries

The thermometer comes with two 1.5V AAA (LR03) alkaline batteries. Replace with two
new AAA (LR03) batteries when the low-battery symbol "
Ensure correct polarity as indicated inside the battery cover. Always replace both batter-
ies at the same time. Remove batteries when the thermometer will not be used for an
extended period of time.

9b. Cleaning and Disinfecting

Use only a soft cloth to clean the surface of the device and LCD with 70% isopropyl
(rubbing) alcohol. Do not use unapproved cleaning or disinfection agents. Use of these
agents may cause damage to components.
Caution: Never use abrasive cleaning agents, thinners, benzene or submerge the ther-
mometer in water.
Caution: Always keep the thermometer within storage temperature range and humidity range.
Caution: Never use water to wash the thermometer lens directly. Slightly moisten a cot-
ton swab or cloth with isopropyl (rubbing) alcohol and gently wipe the surface of the lens.
Caution: Never insert a sharp object into the scanner area or any other open surface
on the thermometer.
Caution: Avoid touching the lens except when cleaning is required.
Note: ADC is not responsible for validating alternative cleaning or disinfection processes
outside of our facility. You should determine the requirements for personnel, equipment,
and materials necessary to achieve your desired disinfection levels and may need to em-
ploy routine monitoring of your disinfection protocols within your facility. After cleaning
or disinfecting, wait at least 10 minutes prior to taking another measurement.
Ambient temperature changes too fast.
1. Power is off.
2. Improper battery installation.
3. The battery is exhausted.
4. Display remains blank.
Wait until the ambient
temperature is stable.
1. Press ON button again.
2. Check the battery polarity.
3. Replace with new batteries.
4. Contact Customer Service.