Cleaning Instruction:
All blood and other body fluids must be thoroughly cleaned from surfaces and
objects before disinfection by the germicidal wipe.
soil.
Disinfection Instruction
:
Unfold a clean wipe and thoroughly wet all the surface of the meter, including the
strip port and other connection port. Unfold a clean wipe and thoroughly wet all the surface of the lancing
device, including cap or AST cap if used. Treated area must remain visibly wet for a full 2 minutes. Let the
devices air dry for 0.5 minute. Do disinfection once per week.
Do:
 Make sure the meter is turned off during cleaning
and disinfection.
 Keep the test strip vials tightly closed when
performing the cleaning and disinfection
procedures because the fumes from the
disinfectant may affect the performance of the strip
 After cleaning or disinfection, please perform the
physical appearance and performance check of
devices.
 Please follow the instruction on page 44 to 46.
Open, unfold and use first germicidal wipe to remove heavy
Do Not:
Get any moisture in the test strip slot.
Spray any cleaning solution directly onto the
meter.
Put the meter under water or liquid.
Pour liquid into the meter.
43