While Recording...
While the UTRIX-16(M) is recording the green
every 4 seconds if none of the configured alarm conditions have been
5
met
.
Alerts
As soon as one of the alarm trigger conditions is met, the red
indicator blinks every 4 seconds. Depending on the alarm
configuration the red indicator may remain on for the rest of the trip,
or the green
indicator may blink again once conditions return
OK
back inside the limits.
If configured, you can reset an alert by pressing the
button. Alerts which are a result of accumulative alarms cannot be
cleared.
Marking a reading with an inspection
mark
When you press the
START
the UTRIX-16(M) will be identified in the downloaded data with an
inspection mark.
If the
Allow stopping with the Stop button
mark will also be registered when you press the
do not complete the process of stopping the recorder.
Stopping the UTRIX-16(M)
Automatically
The UTRIX-16(M) automatically stops recording temperature when
the maximum number of readings specified during configuration has
been reached. Your unit can also be set up to stop when it is inserted
into a USB socket. This option needs to be set up at the factory and
cannot be changed during configuration with LogTag
distributor can supply more information about this option.
Manually
You can configure a UTRIX-16(M) so it can be stopped with the
button. This feature is useful when you take the recorder out
STOP
of a shipment and don't want to falsify the statistics with readings
taken after the shipment completion. The stop function is enabled in
the
Advanced Options
dialogue during configuration. When enabled,
following will stop the unit:
Press and hold the
STOP
l
indicators blink simultaneously.
When the indicators start blinking more rapidly release the
l
button. The recorder will now stop taking readings and cannot be
re-started.
If you change your mind half way during the process, simply keep
l
holding the button until all indicators have stopped blinking.
Instead of stopping, an inspection mark will be recorded against
the next temperature reading taken.
5
The blinking of the green
OK
unit.
Page 8 of 16
indicator blinks
OK
START/Mark
button, the next reading taken by
/Mark
feature is enabled, a
button, but
STOP
®
Analyzer. Your
button. For the next 4 seconds both
indicator can be disabled during configuration of the
Plugging the UTRIX-16(M) into a
USB port
What happens when you plug your UTRIX-16(M) into a USB port
depends on the operating system of the computer, the settings made
during configuration and whether or not LogTag
running.
ALERT
Microsoft Windows
Up to four drivers will now be installed, depending on the UTRIX-16
(M)'s configuration.
All drivers are part of the operating system and will typically not
require administrator privileges for your computer.
1.
Mass Storage Device Driver
2.
LogTag Mass Storage USB Device Driver
These two devices are required so you can access the data files
in the same way as a USB memory stick. These drivers will
not be installed if the UTRIX-16(M) does not generate files.
3.
USB Input Device (HID) Driver
This device is used for communication to LogTag
and its driver will always be installed, even if LogTag
Analyzer is not present on the computer.
4.
USB Composite Device Driver
This driver signals that multiple devices are involved when
plugging in a UTRIX-16(M).
If you have configured the UTRIX-16(M) to
be created every time you plug the recorder into the USB port. While
the file creation takes place, the green
on.
Once finished, the green OK indicator starts blinking every second,
twice in quick succession, to indicate these files can now be accessed.
A new drive or mounted device will appear. The device name will be
created from the serial number of the UTRIX-16(M).
configuration.
If you have disabled file generation, the UTRIX-16(M)'s green OK
indicator will blink once every second.
While a UTRIX-16(M) is plugged into USB, no temperature
readings are taken. The graph will display a gap and the data list
will show --- followed by the # symbol. This also applies when
the recorder is connected to a USB power supply.
OSX and Linux
Typically in these operating systems a new drive will be mounted,
from which you can open the PDF file. You will not be able to
configure the UTRIX-16(M) using either of these operating systems,
unless you use virtualization software such as Fusion or Virtualbox to
create a hosted Windows environment. You need to discuss these
options with your network administrator.
UTRIX-16(M) Product User Guide, Revision 2.0, copyright © 2004-2016,
®
Analyzer is
®
Analyzer
®
generate
files, these will
indicator is permanently
OK
®
LogTag
Recorders
. All rights reserved.