14 Instant help

In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.
Help:
Fault
The displayed weight does
not glow.
The displayed weight is
permanently changing
The weighing result is
obviously incorrect
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
40
Possible cause
• The balance is not switched on.
• Mains power supply interrupted (mains cable
defective).
• Power supply interrupted.
• Rechargeable battery inserted incorrectly or empty
• Draught/air movement
• Table/floor vibrations
• Weighing plate has contact with other objects.
• Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
• The display of the balance is not at zero
• Adjustment is no longer correct.
• Great fluctuations in temperature.
• Warm-up time was ignored.
• Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)