9.4 Linearization (non-verified models only)

Linearity shows the greatest deviation of a weight display on the scale to the value of
the respective test weight according to plus and minus over the entire weighing
range.
If linearity deviation is discovered during a testing instrument control, you can
improve this by means of linearization.
• Carrying out linearization is restricted to specialist staff possessing
well acquainted with the workings of weighing scales.
• The test weights to be used must be adapted to the weighing scale's
specifications; see chapter 2.4 "Testing instruments control".
• Observe stable environmental conditions. Stabilisation requires a
certain warm-up time.
• After successful linearization you will have to carry out calibration;
see chapter 2.4 "Testing instruments control"
Tab. 1: Adjustment points
Adjustment
weight
1.
2.
3.
4.
GAB-N-BA-e-1817
GAB 6K0.05N
0 kg
2 kg
4 kg
6 kg
GAB 12K0.1N
0 kg
4 kg
8 kg
12 kg
GAB 30K0.2N
0 kg
10 kg
20 kg
30 kg