Deepace KC908 Manual page 91

Real-time spectrum analyzer 6g
Table of Contents

Advertisement

Demodulation parameters are displayed on the information bar on the
middle on the screen.
(14) Adjust the antenna , find out the position and direction where field
strength maximizes and watch the reading on the screen. For screenshots,
press SHIFT+1.
(15) Accumulating function, which is an optional component, can be used
if the standard of the test demands. The entrance locates at mode main menu.
The "accumulating method" setting in this function is comprised of peak and
average. This detection is executed once again after the detection of the
machine itself. For instance, average detecting method has been selected in
MODU menu, yet peak accumulating method is set in accumulating menu,
then the machine will pick the peak value among several average values. The
accumulating time can also be set, such as a typical value of 360 seconds in
electromagnetic radiation standard.
(16) Provided that these mention settings are freqeuently used, they
should be stored as a channel. To do that, press VFO in mode main menu,