7. Open the protective slider on the calibration holder to access the white calibration tile.
8. Position the i1Pro in the calibration holder with the aperture over the white calibration tile.
9. After the i1Pro is positioned in the holder, tap the Calibrate i1 button on the screen to perform a
white calibration.
10. After white calibration is complete, remove the i1Pro from the holder and close the slider in the
holder over the white tile.
11. Attach the ambient light measuring head on the limiting aperture of the i1Pro. The arrow on the
protective cover must point to the "O" unlock marking on the i1Pro.
12. Rotate the protective cover with ambient light measuring head clockwise until the arrow points to
the "
" locked marking on the i1Pro.
13. Pull the protective cover from the ambient light measuring head.
V i r t u a l L i g h t B o o t h
White calibration tile
Protective slider
21