This supplement contains information necessary to ensure the accuracy of
the above manual. Enter the corrections in the manual if either one of the
following conditions exist:
1.
The revision letter stamped on the indicated PCA is equal to or higher
than that given with each change.
2.
No revision letter is indicated at the beginning of the change.
© 2001 Fluke Corporation. All rights reserved. Printed in U.S.A.
Manual Supplement
i1000s Calibration
1574933
May 2001
Supplement Issue:
Issue Date:
Page Count:
1
7/01
2

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the i1000s and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Fluke i1000s

  • Page 1 The revision letter stamped on the indicated PCA is equal to or higher than that given with each change. No revision letter is indicated at the beginning of the change. © 2001 Fluke Corporation. All rights reserved. Printed in U.S.A.
  • Page 2 Calibration Manual Supplement Change #1 On page 8, Table 4, under Recommended Model, Replace: Digital Multimeter Hp 3458A or equivalent With: Digital Multimeter Agilant 3458A or equivalent Delete the following from Table 4, AC/DC Shunt IET Labs-DCCS-200 0.1e - 0.1% non-inductive...
  • Page 3 Output Test 1. Connect the equipment as shown in Figure 5. 2. Set the calibrator output and coil for each step of Table 5. 3. Verify that the voltmeter reading is within the “i1000S mV Output” limits of Table 5. 7/01...