Stop Measurement (Idle); Ping User Data - Leister axetris LGD Compact Series Operating Instructions Manual

Laser gas detection oem module
Table of Contents

Advertisement

Stop measurement (Idle)

Idle
Send the
-command to stop the current measurement. This allows accessing (
Command in Hex
Command in ASCII
As a confirmation, the device responds with an
and waits for further commands. It won't send any further data or packages.

Ping user data

Ping
With the
command user data can be asked from the device. Please note that the device needs to be in
Idle
-mode to have access to the user data.
Command in Hex
Command in ASCII
Ping
The device transmits a
execution):
Response
Table 14
from the device with description of the bytes
Byte
Possible HEX
1 – 4
7B 50 50 00
5 – 6
00 00
7 – 10
00 00 40 40
11 – 14
CD CC 8C 3F
15 – 18
00 00 00 00
19 – 22
00 00 00 00
23 – 26
00 00 00 00
27 – 30
00 00 C8 42
31 – 34
00 00 00 00
35 – 38
0A D7 23 3C
39 – 42
00 00 20 41
43 – 46
00 80 BB 44
47 – 50
33 33 73 3F
51 – 54
00 00 00 00
55 – 58
00 00 00 00
59 – 62
00 00 00 00
63 – 66
00 00 00 00
67 – 70
00 00 00 00
Axetris AG
LGD Compact Operating & Integration Instructions; Rev. D
Idle
(without any data). The device now is in
-packet with the following content (content can vary depending on product
Start of packet with size (80 bytes)
status code bytes
Concentration gas 1 Offset
Concentration gas 1 Span
Reserved variable for future use (dummy)
Reserved variable for future use (dummy)
Analog output: Min concentration
Analog output: Max concentration
Analog output: Offset
Analog output: Span
Integration time
Concentration gas 2 Offset
Concentration gas 2 Span
Diagnostic Mode (1 = active / 0 = inactive)
For further details about Diagnostic mode, contact Axetris AG.
Reserved variable for future use (dummy)
Reserved variable for future use (dummy)
Reserved variable for future use (dummy)
Reserved variable for future use (dummy)
7B 49 08 00 00 EA
4A
7D
{ I . . . .
J
}
7B 50 08 00 10 00
1D
{ P . . 0 .
.
}
Description
Ping
) the user parameters.
Idle
-mode (standby)
7D
(here 3)
(here 1.1)
(here: 0)
(here: 100)
(here: 0)
(here 0.01)
(here: 10)
(here 1500)
(here 0.95)
(here 0)
31/46

Advertisement

Table of Contents
loading

This manual is also suitable for:

Axetris lgd compact a

Table of Contents