D-Sub sensor module J2 (15pol, 3 rows)

The sensor module is connected to the electronic housing via a three row D-sub connector.
Figure 12 shows the pin assignment.
Table 4 Connector assignment 15pol D-Sub (3 rows)
Pin
Measurement
11
Signal
12
Laser control
1
9
5
8
6
2
3
10
7
4
Temperature
13
measurement
14
Not used
15
Shield
Case
* Same potential
Axetris AG
LGD Compact Operating & Integration Instructions; Rev. D
Name
Description
PD_I
Photodiode
PD_GND*
Laser_Drive
Laser diode
Laser_GND
Laser_Supply3V3
Laser_U
diff
Laser_ID
Laser_Thermistor
Thermal resistance for laser temperature measurement
Laser_GND*
Laser_TEC+
Peltier element for laser temperature stabilization
Laser_TEC-
Laser_PT1K
Thermal resistance for optical head temperature
measurement
Gascell_PT1K
Thermal resistance for gas temperature measurement
Gascell_GND*
Reserve
n/a
SHIELD
Connected from D-Sub casing to cable shield
Figure 12 Sensor module connector
assignment