Figure 4 Dimensions of the main board
The electronic main board should be connected to
the sensor module just with the original cable
delivered (recommendations for an own cable design
can be seen in chapter 3.1.4).
The right position for connecting the cable to the
sensor module are marked on the cable connector
as well as on the sensor module. To fix the cable, the
wheel must be screwed counter-clockwise.

2.3 Electrical Interfaces

The standard LGD interface is a RS232 connection. Connection to a USB port is possible via an RS232/USB
adapter supplied with the system. Analog 4-20 mA or 0-5 V interfaces will be factory calibrated upon customer
request. More details can be also found at chapter 1.2.4.
To connect the system via a USB port please only use our tested RS232/USB adapter that is provided
with the system.
Please avoid starting-up or restarting a computer when the LGD is connected to the serial interface or
through a USB port.
Axetris AG
LGD Compact Operating & Integration Instructions; Rev. D
Figure 5 Dimensions of the interface board
Figure 6 Correct connection of cable and sensor
module
15/46