Performance Verification
3–4
The fuse location
Figure 3–1:
5. Using needle-nose pliers, gently pull out the fuse.
CAUTION. To avoid the risk of fire, replace a damaged fuse only with a fuse of
the specified type and rating.
6. Replace the fuse with a 250 VAC 2 A model 3AG Med-Blo fuse.
7. Seat all ribbon cables by pressing firmly on all the connectors.
8. Slide the top panel back onto the cabinet.
9. Replace the two screws on the rear panel.
10. Plug the Preselector back in.
The Preselector is now ready for use.
2706 RF Preselector