Table of Contents Preface ......................Clear and Sanitize Procedures..................Memory Devices....................Data Export Devices................... Troubleshooting..................... How to Clear or Sanitize a Nonfunctional Instrument............. How to Recover from Clearing or Removing the Instrument’s Memory ........ 1741C Declassification and Security...
Page 4
Table of Contents 1741C Declassification and Security...
Preface This document helps customers with data security concerns to sanitize or remove memory devices from the 1741C Analog Dual-Standard Waveform Monitors. These products have data storage (memory) devices and data output devices (USB ports). These instructions tell how to clear or sanitize the memory devices and disable the data output devices.
Page 6
Scrub. This is equivalent to sanitize. User-modifiable. The user can write to the memory device during normal instrument operation, using the instrument interface or remote control. Volatile memory. Data is lost when the instrument is powered off. 1741C Declassification and Security...
flash memory instrument for at least 20 service manual for seconds removal instructions. Store the removed board in a secure area or destroy it. Contact Tektronix for board replacement and adjustment 1741C Declassification and Security...
Page 8
20 service manual for memory seconds removal instructions. Store the removed board in a secure area or destroy it. Contact Tektronix for board replacement and adjustment 1741C Declassification and Security...
Page 9
During normal instrument operation. Before you remove any module or board for the purpose of sanitizing the instrument, please contact your local Tektronix Customer Service Center to check on the availability and cost of replacement parts. Table 2: Nonvolatile Memory Devices...
Page 10
During normal instrument operation. Before you remove any module or board for the purpose of sanitizing the instrument, please contact your local Tektronix Customer Service Center to check on the availability and cost of replacement parts. Clear Presets Procedure 1.
Page 11
3. Navigate to IP Address. Then press > to enter the edit mode. Enter “000.000.000.000” for the IP address. 4. Repeat Step 3 for the Subnet Mask and Gateway Address. 5. Press the CONFIG button to exit the configuration menu. 1741C Declassification and Security...
3. Press the CONFIG button to exit the configuration menu. Enable Ethernet Access To enable Ethernet access, use the same procedures you would use to disable these devices, but select On to enable each device. Procedure 1741C Declassification and Security...
How to Clear or Sanitize a Nonfunctional Instrument To sanitize a nonfunctional instrument, remove the Main board and return the instrument to Tektronix for installation of a new Main board. This procedure does not clear calibration constants stored on the Audio, Eye, and Composite boards.