Tektronix 1741C Declassification And Security Instructions
Tektronix 1741C Declassification And Security Instructions

Tektronix 1741C Declassification And Security Instructions

Analog dual-standard waveform monitor
Hide thumbs Also See for 1741C:

Advertisement

1741C
xx
Analog Dual-Standard Waveform Monitor
Declassification and Security
ZZZ
Instructions
www.tektronix.com
077-0153-00

Advertisement

loading
Need help?

Need help?

Do you have a question about the 1741C and is the answer not in the manual?

Questions and answers

Summary of Contents for Tektronix 1741C

  • Page 1 1741C Analog Dual-Standard Waveform Monitor Declassification and Security Instructions www.tektronix.com 077-0153-00...
  • Page 2 Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material.
  • Page 3: Table Of Contents

    Table of Contents Preface ......................Clear and Sanitize Procedures..................Memory Devices....................Data Export Devices................... Troubleshooting..................... How to Clear or Sanitize a Nonfunctional Instrument............. How to Recover from Clearing or Removing the Instrument’s Memory ........ 1741C Declassification and Security...
  • Page 4 Table of Contents 1741C Declassification and Security...
  • Page 5: Preface

    Preface This document helps customers with data security concerns to sanitize or remove memory devices from the 1741C Analog Dual-Standard Waveform Monitors. These products have data storage (memory) devices and data output devices (USB ports). These instructions tell how to clear or sanitize the memory devices and disable the data output devices.
  • Page 6 Scrub. This is equivalent to sanitize. User-modifiable. The user can write to the memory device during normal instrument operation, using the instrument interface or remote control. Volatile memory. Data is lost when the instrument is powered off. 1741C Declassification and Security...
  • Page 7: Clear And Sanitize Procedures

    flash memory instrument for at least 20 service manual for seconds removal instructions. Store the removed board in a secure area or destroy it. Contact Tektronix for board replacement and adjustment 1741C Declassification and Security...
  • Page 8 20 service manual for memory seconds removal instructions. Store the removed board in a secure area or destroy it. Contact Tektronix for board replacement and adjustment 1741C Declassification and Security...
  • Page 9 During normal instrument operation. Before you remove any module or board for the purpose of sanitizing the instrument, please contact your local Tektronix Customer Service Center to check on the availability and cost of replacement parts. Table 2: Nonvolatile Memory Devices...
  • Page 10 During normal instrument operation. Before you remove any module or board for the purpose of sanitizing the instrument, please contact your local Tektronix Customer Service Center to check on the availability and cost of replacement parts. Clear Presets Procedure 1.
  • Page 11 3. Navigate to IP Address. Then press > to enter the edit mode. Enter “000.000.000.000” for the IP address. 4. Repeat Step 3 for the Subnet Mask and Gateway Address. 5. Press the CONFIG button to exit the configuration menu. 1741C Declassification and Security...
  • Page 12: Data Export Devices

    3. Press the CONFIG button to exit the configuration menu. Enable Ethernet Access To enable Ethernet access, use the same procedures you would use to disable these devices, but select On to enable each device. Procedure 1741C Declassification and Security...
  • Page 13: Troubleshooting

    How to Clear or Sanitize a Nonfunctional Instrument To sanitize a nonfunctional instrument, remove the Main board and return the instrument to Tektronix for installation of a new Main board. This procedure does not clear calibration constants stored on the Audio, Eye, and Composite boards.