Device status (DIAG_STATUS)

Meaning of "Allgemeiner Status" as per NAMUR NE107:
F: Error (Failure)
C: Functional test (Function Check)
O: Outside of specification (Out of Specification)
M: Maintenance required (Maintenance Required)
Bit
DIAG_STATUS
31
F: Failure
30
C: Function Check
29
O: Out of specification
28
M: Maintenance required
46
PV
Cause
Gas reading
– (Bit not used)
Gas reading
– (Bit not used)
Gas reading
Calibration interval expired.
Gas detector is operated outside the specified
temperature range.
Gas reading
Gas reading outside the measuring range.
Gas reading
– (Bit not used)
Gas reading
– (No warning, no error)
General status
Instructions for use
Cause
®
|
Dräger Polytron
6100 EC WL, Dräger Polytron
Remedy
Calibrate the gas detector.
Use weather protection kit (accessory) when
exposed to sunlight.
A Calibrate the gas detector.
B Change the sensor.