Checking The Original Dimensions; Rechargeable Battery Operation - KERN HFD Service Manual

Electronic crane scales
Table of Contents

Advertisement

6.2 Checking the original dimensions

Enter the original dimensions shown on the production data sheet in the grey fields of the
checklist chap. Fehler! Verweisquelle konnte nicht gefunden werden.. in the HFD user manual
Check original dimensions of crane balance; for implementation see chap.15.2 "Regular
Maintenance" in the HFD user manual
Enter all data (date, tester, results) in the first line under "Inspection before first use" in the
checklist (see chapter Fehler! Verweisquelle konnte nicht gefunden werden. in the HFD user
manual)
If the dimensions of your first safety inspection do not match those of
KERN, the balance must not be put into operation. In this case please
contact a service partner authorised by Messrs. KERN.
CAUTION

6.3 Rechargeable battery operation

CAUTION
Before the first use, the rechargeable battery should be charged by connecting it to the mains
power cable for at least 24 hours. The operating time of the rechargeable battery is approx. 60
hours.
The capacity of the rechargeable battery will soon be exhausted, when the display begins to flicker
"lo_bat" appears, the balance will remain operable for approx. 30 minutes more, then it switches
off automatically. Connect the power cable as soon as possible to load the rechargeable battery.
The LED display over
red:
Voltage has dropped below prescribed minimum.
green:
Rechargeable battery completely reloaded
yellow:
Charging storage battery
To save the battery, the balance switches automatically off without weighing (time can be
selected, see chap. Fehler! Verweisquelle konnte nicht gefunden werden.. Function "F9 of" in
the HFD user manual)
When the crane scales are out of operation for a longer period, remove the rechargeable battery.
HFD-SH-e-1810
Material damage on crane scales
Only use the delivered mains adapter.
Make sure that the mains adapter, the cable and the mains
plug are in a perfect condition.
Do not use the crane scales during the loading process.
provides information about the battery's charging status.
. If
12

Advertisement

Table of Contents
loading

Table of Contents