Basic Information; Introdution - KERN HFD Service Manual

Electronic crane scales
Table of Contents

Advertisement

1 Basic Information

The device must be repaired only by trained specialist staff or personnel with professional
formation (such as a repair-specialist accredited by law concerning verification). The service
manual is obligatory for repair work. After repair, original conditions of the device have to be
restored. Only original spare parts should be used.
Instructions about conformity-evaluated scales:
Repair must be carried only at 100% compliance with the type approval. A violation of this
specification will result in a loss of the type approval! After successful repair the balance will have
to be reverified before it can be used again in a statutorily regulated field.
Detailed instructions about conformity-evaluated scales:
Repair must be carried only at 100% compliance with the type approval. A violation of this
specification will result in a loss of the type approval!
After successful repair the balance will have to be reverified before it can be used again in a
statutorily regulated field.

2 Introdution

This service manual covers the HFD series and is edited for the authorized servicing personnel.
Note all rights are reserved. Copying any part of this manual is prohibited without our permission.
In this lineup it is an eco product, Therefore, it is not intended to represent the repair manual in
detail, since the construction of the balance is very simple. It is therefore only referring to the list
of related to disposal spare parts.