This is the sequence used by the PMM Emission Suite when a scantable is run, in
7.7 Sweep Mode
the "conducted" range (Band B from 0.15 to 30 MHz).
commands
sequence
example
Command sent
Reply received
#?IDN*
IDN=ER8000/00-FW - 1.00
11/06/20
#?S/N*
S/N=000WE20304
#?CRA*
CRA=OK
#SCFA -1*
CFA=OK (OFF)
#S3PRC*
3PR=OK
#SSFD
150000;30000000;5000;P;2;6;10;OFF;OFF;0*
Binary data reply
#?UHT*
UHT=2.5ms
7-20
Remote control
Example
Queries information about model, release and date of the firmware
It is a ER8000/00 with its FW
It is used to check congruencies, frequency limits, and all the related
characteristics
Queries the serial number
Serial number
Queries if RMS-AVG optional detector is active
RMS-AVG active
Correction factors, if present, deactivated
Important. Could be previously stored and activated.
Sets the unit in the bands A and B for conducted
ack
Sweep
Measurements data
Queries the hold time
Applied hold time
Description