This setting command sets the Margin to be used with Smart Detector function. The
SLIM n
argument (n) in the range of -20 to 20.
The reply is LIM =OK which acknowledges the command has been granted or LIM
=SERR if the command has been ignored.
Example: #SLIM 2* Sets the threshold of smart detector to 2 dB below the limit.
SLIW
This setting command is used for setting the custom limit frequency by frequency. It
consists of three fields as follows:
n,freq;lev
•
•
•
The procedure to set a custom limit on PMMER8000 is the following:
Here is an example how to make a custom limit on PMMER8000:
Note that all the higher indexes in the limit are cleared. Therefore, a limit must be made
in upwards order otherwise all points, having a higher index previously written, are
cleared.
It is responsibility of the user to make sure the data sent are correct and coherent.
The reply is SLIW =OK which acknowledges the command has been granted or LIW
=SERR if the command has been ignored.
SLSN n
This setting command #SLSN n* sets the input for Band B according to the argument 'n'
as follows:
•
•
•
'n' higher than '2' is ignored and SERR message sent back
The reply is LSN =OK which acknowledges the command has been granted or LSN
=SERR if the command has been ignored.
Example: #SLSN 0* switches the input from 'N' connector
Example: #SLSN 2* switches the input from ER8000 built-in LISN Line-2
This setting command sets the tuning frequency for manual mode. The string (f) can be
SMAF f
in exponential form and should be expressed in Hz. The reply is MAF=OK which
acknowledges the command has been granted or MAF=SERR if the command has
been ignored.
When the command is granted the ER8000 goes to manual mode automatically
Example: #SMAF150e3* Sets the manual frequency to 150kHz
7-12
Remote control
n Is the index of frequency being written and it should range from 0 to 15
freq Is the frequency, expressed in Hz (exponential notation allowed), being
written
lev Is the level of the limit and is expressed in dBµV
•
Send as many commands as the frequency points of limit are in
upwards order.
•
Terminate (and make it active) by sending the command SLIE
•
#SLIW 0, 150e3; 66 * Send the first line 150kHz 66 dBµV
•
#SLIW 1, 500e3; 56 * Send the second line 500kHz 56 dBµV
•
#SLIW 2, 5e6; 56 * Send the third line 5MHz 56 dBµV
•
#SLIW 3, 5e6; 60 * Send the fourth line 5MHz 60 dBµV
•
#SLIW 4, 30e6; 60 * Send the fifth line 30MHz 60 dBµV
•
#SLIE Custom CISPR 14-1 * Send the name and make the limit active
0 Input on N connector
1 Input on ER8000 internal LISN L1
2 Input on ER8000 internal LISN L2