L3Harris NARDA PMM ER8000 User Manual page 56

Emi receiver
Table of Contents

Advertisement

Following are Commands to control and set various operating modes of PMM
7.6 PMM ER8000
ER8000, that can be distinguished in two main groups:
COMMANDs
7.6.1 Query
COMMANDs
?3PR
This query command #?3PR * sends back a string expressing whether the
apparatus is a PMM ER8000 and in which mode. The reply is either:
?AAT
This query command #?AAT * sends back a string expressing the Attenuator status
for Analyzer mode. The reply is made of 2 fields:
Example of reply: AAT =AUTO; 20 which means that the Attenuator is now
switched to 20dB automatically by ER8000
?ACE
This query command #?ACE* sends back a string expressing the Center
frequency, in exponential notation, for analyzer mode. Unit is fixed MHz.
Example of reply: ACE = 1.500000e+07 which means that the Center frequency in
Analyzer mode is 15MHz
?ADT
This query command #?ADT * sends back a string expressing the detector used for
Analyzer mode. The reply cab be:
?AHT
This query command #?AHT* sends back a string expressing the HOLD TIME for
Analyzer mode. Unit is fixed ms.
Example of reply: AHT= 1000 ms which means that the hold time in Analyzer mode
is 1000ms (or 1s)
?AOP
This query command #?AOP* sends back a string expressing the Stop frequency,
in exponential notation, for analyzer mode. Unit is fixed MHz.
Example of reply: AOP = 3.000000e+07 which means that the Stop frequency in
Analyzer mode is 30MHz
?ART
This query command #?ART* sends back a string expressing the Start frequency,
in exponential notation, for analyzer mode. Unit is fixed MHz.
Example of reply: ART = 1.500000e+07 which means that the start frequency in
Analyzer mode is 15MHz
?ASP
This query command #?ASP* sends back a string expressing the Spanfrequency,
in exponential notation, for analyzer mode. Unit is fixed MHz.
Example of reply: ASP = 3.000000e+07 which means that the Span frequency in
Analyzer mode is 30MHz
7-4
Remote control
Query Commands
Setting Commands
3PR =CON (The PMM ER8000 is ready for conducted frequency range )
3PR =RAD (The PMM ER8000 is ready for radiated frequency range )
3PR=NUL (The PMM ER8000 is not ready)
AAT =AUTO (ER8000 automatically selects the most suitable attenuator to
get best dynamic range) or AAT =MAN ( operator selects manually the
attenuator)
Value (in dB) of current attenuator
ADT =Peak
ADT =Avg
ADT =Rms
Description

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents