L3Harris NARDA PMM ER8000 User Manual page 65

Emi receiver
Table of Contents

Advertisement

This setting command puts the ER8000 in manual mode. The reply is always
SMAN
MAN=OK which acknowledges the command has been granted.
The ER8000 goes to manual mode.
Example: #SMAN *
This setting command puts the ER8000 in manual mode and sets the default value
SMANP
to the following settings:
The reply is always MANP=OK which acknowledges the command has been
granted.
The ER8000 goes to manual mode.
Example: #SMANP*
This setting command sets Attenuator, for Manual Mode, to the value indicated by
SMAT a
(a) which can be 0 to 45. Automatic Attenuator feature is disabled. If (a) is a negative
figure then the Automatic Attenuator feature is turned on. The reply is MAT=OK
which acknowledges the command has been granted or MAT =SERR if the
command has been ignored.
When the command is granted the ER8000 goes to manual mode automatically.
Example: #SMAT 15* (set attenuator to 15dB and disable Automatic Attenuator
feature)
Example: #SMAT -1* (Enable Automatic Attenuator feature)
SMHT h
This setting command sets hold time for manual mode. The string (h) should be
expressed in ms. The reply is MHT=OK which acknowledges the command has
been granted or MHT =SERR if the command has been ignored.
When the command is granted the ER8000 goes to manual mode automatically.
Example: #SMHT 1500* Sets the hold time, for manual mode, to 1500 ms (or 1.5s)
This setting command sets RBW. The argument (b) should be a string representing
SRBW b
the RBW as follows:
The reply is RBW =OK which acknowledges the command has been granted or
RBW =SERR if the command has been ignored.
Example: #SRBW 3* sets the RBW to 30kHz and disables Auto RBW mode
Example: #SRBW 0* turn the RBW mode in Automatic
RBW Automatic
Attenuator Automatic
MinAtt=10 dB
Preselector On
Preamplifier OFF
Hold Time =1000ms
Pulse Limiter = OFF
0 RBW is Automatic so that the ER8000 automatically selects
o
the suitable RBW according to the tuned frequency
1 300kHz
o
2 100kHz
o
3 30kHz
o
4 10kHz
o
5 3kHz
o
6 9kHz (Cispr 16-1)
o
7 200Hz (Cispr 16-1)
o
8 1kHz
o
Only for C, D band o > 30 MHz
o
10 120kHz (Cispr 16-1) (Opt. 01)
9 1MHz (Opt. 01)
Remote control
7-13

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents