Relay Contacts; Delay Adjust; Milliamp Output (+Ma / -Ma); Considerations For The Flare Monitor (Fm) - Advanced Energy LumaSense Technologies E2T QUASAR Series Operation Manual

Flare monitors
Table of Contents

Advertisement

b. Verify that noise amplification has not returned. If the noise has returned, proceed
to step 5.
c.
If noise has not returned, the procedure is completed. (Proceed to 6.1.8, Relay
Contacts.)
5. Reduce the delay time and repeat the above procedure.
a. If noise amplification continues, repeat the setup adjustment procedures starting at
Section 6.1.4, System Gain/Climate Compensation.
b. If problems persist, contact the factory for further technical assistance. (Refer to
Section 1.5 for more information).
6.1.8

Relay Contacts

Check the relay contacts at the instrument's Terminal Output board. N.O. and N.C. contacts are
both available simultaneously. Verify that the desired polarity is obtained during an alarm
condition.
Note: Make note of the contact position when the instrument power is interrupted
or shut-off. The relay is energized when the green LED is lit. It is not energized during
an alarm or power loss.
6.1.9

Delay Adjust

The Delay Adjust (Item 9) can be configured from 2 seconds (completely counterclockwise) to
approximately 2 minutes (completely clockwise). The Delay Pot postpones the Alarm condition
once the set-point value is exceeded. This can prevent false alarms from temporary loss of signal
due to wind.

6.1.10 Milliamp Output (+mA / -mA)

Check the milliamp output signal at the instrument's Terminal Output board. The signal should
be approximately 4mA when the flame is present and the analog meter reads full-scale. The
signal should read approximately 20mA when flame is lost, with analog meter under .35 of
scale.
Milliamp checks must be accomplished with a Milliamp meter placed in series with the control
room loop or parallel across the mA+ and mA- terminals when no other connections are on the
terminals.
1. Replace the top cover and the sixteen M8 x 1.25 socket head screws, torque to 30in/lbs.
2. Supply positive air flow and pressure to the air connections.
3. Verify that the temperature of the QUASAR M8100-EXP housing will not exceed 120 ° F
(50 °C). If cooling is required, refer to Chapter 3, Mechanical Installation.

6.2 Considerations for the Flare Monitor (FM)

In the case where it is not possible to eliminate the Pilot flame completely from the instrument's
field of view, it is necessary to verify that a "Flame-out" alarm will occur as expected. It is
recommended that you complete the following procedure to confirm proper operation. Be sure
to complete the start-up checklist procedure first. Refer to Section 6.1 of this Chapter.
1. Ensure that the Pilot flame is present and that the Flare is in the condition where the
operator wishes to see an Alarm from the instrument.
E
T QUASAR Flare Monitors Operation Manual
2
Operating the M8100FM-EXP • 39

Advertisement

Table of Contents
loading

Related Products for Advanced Energy LumaSense Technologies E2T QUASAR Series

Table of Contents