4.3 Fusing

4.3.1

Mains

Two fuses, F1 and F2, are each mounted within fuse holders on the Terminal Output PC Board,
located directly under TB1.
F1 is used for all power configurations.
F2 is used only in 220 / 230 V AC applications.
Note: The factory installed F2 bypass jumper J1, is hardwired for 110 V AC and
24 V DC systems. Refer to Appendix A for Power and Fuse requirements.
4.3.2

Heater

Two fuses, F3 and F4, are each mounted within fuse holders on the Terminal Output PC Board
located near the Heater connector SKT1.
F3 is used for all power configurations.
F4 is used only in 220 / 230 V AC applications.
Note: The factory installed F2 bypass jumper J2, is hardwired for 110 V AC and
24 V DC systems. Refer to Appendix A for power and fuse requirements and
Appendix E, drawing 613-816, Service Connections.

4.4 Connection for mA Output

The mA current loop is a self-powered output for recorders or controllers requiring current loop
feedback. The QUASAR M8100-EXP supplies its own 4 ... 20mA current to a load. The 4 ... 20mA
output is isolated from the Power ground. Loads up to a maximum of 400 ohms may be
connected in SERIES. Once connected, voltage measured between mA outputs and Chassis GND
(common mode) should not to exceed 20 V DC.
A shield is recommended, however do not connect the shield to the instrument's GND, connect
it only to the Chassis GND.
Caution: Refer to Chapter 10, Safety Assurances and Precautions before
connecting the 4 ... 20mA loop to the terminal.
Connect the mA lines to the corresponding (MA-) and (MA+) outputs on TB2.
Positive mA to mA+
Negative mA to mA-
E
T QUASAR Flare Monitors Operation Manual
2
Electrical Installation • 26