8.7 Sizing tables
Sizing table (DN
/
" to DN 78")
1
12
The tables show the relationship between flow velocity v, flow quantity Q and sensor dimension
DN.
Guidelines for selection of sensor
Min. measuring range: 0 to 0.25 m/s (0 to 0.8 ft/s)
Max. measuring range: 0 to 10 m/s (0 to ft/s)
Normally the sensor size is selected so that the nominal flow velocity v lies within the measuring
range 1 to 3 m/s (1 to 15 ft/s).