Figure 2-2
MAG 5100 W (7ME6520) MI-001 is verified and labeled at a given Q3 and Q3/Q4 = 1.25 and
Q2/Q1 = 1.6 measuring ranges see appendix Measuring ranges (Page 67)
2.6.2

PTB K7.2

MAG 5000/6000 CT together with MAG 5100W (7ME652) are approved for PTB K7.2 under the
following installation conditions:
• DN 50 to 300 mm (2" to 12")
• Horizontal installation
• Compact or remote with max. 10 m (33 ft.) cable
Other restrictions may apply (see certificate 22.76/10.02)
Measuring range according to EN1434 (2007), see appendix Measuring range according to
EN1434 (2007) (Page 69)
SITRANS FM MAG 5100 W
Operating Instructions, 01/2020, A5E03063678-AC
Nameplate MAG 6000, 5100 W
Safety notes
2.6 Custody transfer
17