ACM 150
Temperature and protection
Install indoors in a heated and air conditioned environment. The recommended
temperature range is 60 to 85 °F (15 to 30 °C). The standard cabinet is designed to
meet an IP 20 rating.
Electrical area
The ACM 150 monitor is designed to be located in a non-hazardous or general
purpose electrical area. Refer to Technical Specifications on page 97.
Utility requirements
The ACM 150 monitor requires the following utilities:
Line power: 115V 60Hz 13 Amps or 230V 50Hz 6 Amps (setup at factory)
Normal idle power consumption: ~8 amps @ 115 VAC
Average power consumption: ~14 amps @ 115 VAC
Peak power consumption: ~20 amps @ 115 VAC
Clean, Dry Air (CDA): 90 psig, 12 SCFM, 3/8 inch OD tube
Nitrogen: 5 psig, ~8 liters/minute, 1/4 inch OD tube
Exhaust: 15 cfm, 1-1/2 inch OD tube
Main supply voltage: ±10% of nominal voltage
Overvoltage: Category II
 