Software Servicing:
May be used to analyze and control the database on the ACM 150 monitor and is
designed for our Fieldservice Engineers.
Manual Mode Op:
When selected, the manual mode menus appear on the right.
Mode Control:
Users with an access code may change the mode from automatic to manual or
frontpanel.
View Valve:
Same screen as View Valve Cycles, but the user has manual control over the
valve cycling.
Sample Valves:
The port valves 1-40 may be manually controlled in this menu.
Relays:
The 40 programmable relays may be manually activated.
Alarm Simulation:
Allows the user to simulate gas concentrations at the Modbus / TCP
interface.
Service log
Reviewing service events
The 'Service log' key pulls up the menu shown in
currently stored in memory are displayed starting with the most recent. This list may
be limited to a shorter time period, e.g. 1 day or 2 days, by selecting the number of
days using the arrow keys and then pressing "Limit List".
These are same incidents listed in the Service Status mode, except you can view
incidents that have been cleared and no longer appear in the Service Status mode.
NOTE
To troubleshoot a current problem, refer to
instructions provided. Alternately, you can select 'View Instructions' and pull up
'Online Help', which is not currently available.
 