Imb-Pm1; Wiring Of Imbus Switching Modules 6.1 Imb-Io4 - IBR IMB-im2 Instruction Manual

Imbus measuring bus
Table of Contents

Advertisement

Instruction Manual
In addition to the IMB-sm1/2/4 modules connection cables / -adapters for the connection of the par-
-
ticular gauges are offered.
2. Configuration :
-
Open Setup-Window of IMB_Test programme.
-
Press Setup-Button behind the input ( Column Gauge / Sensor ) :
-
Selection of the gauge

5.8 IMB-pm1

1. Short description :
-
The IMB-pm1 modules allow the connection of gauges with „old" parallel interface.
2. Configuration :
-
see 5.7.2.
6
Wiring of IMBus switching modules
6.1 IMB-io4
1. Short description :
The IMBus input and output modules are all galvanically ( 2kV ) separated.
-
-
The inputs are compatible to PLC optocoupler inputs and work in a wide voltage area.
The outputs are ESD proofed, short circuit proofed and contain a high driver power.
-
-
Each in-/output contains a state LED.
The connections are done by plugable terminal strips.
-
2. Technical data :
Connection of 2-wire-BEROS is possible
IMBus
Nominal input voltage
Input voltage signal "High"
Input voltage signal "Low"
Input current signal "High"
Isolation voltage
24
Electrical characteristics "Input"
24 V DC
13 V ... 30 V
-30 V ... + 5 V
typ. 7 mA
max. current 1,5 mA
2000 V