Instruction Manual
5. Calibration with 2 masters ( = Button „Calibration" ):
-
The menu „Calibration" allows the calibration of the measuring inputs with 2 masters.
The calling of the menu is only necessary in some special cases, because the IMB-im1/2/4/8 modules
-
are calibrated at the manufacturer with special reference calibration equipment.
-
Input passwort „scal".
-
Input the master values.
-
Press the „Cal"-Button of the particular measuring input.
After that the software requests to insert the first master and then to insert the second master. After
-
collection of the measuring values of both masters the software calculates automatically the gain and
offset and stores both values inside the module.

5.2 IMB-dm1/2/4

1. Short description :
-
The IMB-dm1/2/4 modules allow the connection of incremental measuring systems with 1Vpp output
or together with a special adapter cable the connection of measuring systems with 11µA output.
-
The pining fulfils the Heidenhain Standard.
2. Configuration :
Open Setup-Window of IMB_Test programme.
-
-
Press Setup-Button behind the input ( Column Gauge / Sensor ) :
IMBus
22