2. Configuration :
Open Setup-Window of IMB_Test programme.
-
-
Press Setup-Button behind the input ( Column Gauge / Sensor ) :
-
Measuring direction :
Standard : On pressing in the probe the measuring value becomes bigger.
Inverse :
-
Measuring step :
0.0001 :
0.00025 :
!!! Please request this setting from the IMBus manufacturer or distributor.
-
Digital Filter :
( for filtering i.e. vibrations on the machine )
off :
Level 1 :
....
Level 8 :
3. Mechanical adjustment of inductive probes ( = Button „Mech. Adjust." ):
-
The menu „Mech. Adj." allows the mechanical adjustment of the probes in the fixture.
-
The inductive probes contain their smallest linearity error near their mechanical zero point and so the
probes must be clamped near their mechanical zero point.
-
Insert master into the fixture.
Insert probe into the fixture and move it in position, that the column display of the particular probe is
-
green.
Fix screw for clamping the probes in this position. ( Note : The screw must not be tightened too strong,
-
otherwise the probe can be damaged ).
4. Zero adjustment with one master ( = Button „Zeroadjust." ):
-
The menu „Zeroadj." allows to set the probes on the master to zero.
-
This menu is unnecessary together with most software packets ( like ComGage, IBREXDLL, ... ) be-
cause most application software packets have an own zero-adjustment possibility.
Insert master into the fixture.
-
-
Press button „Adj" of particular probe.
IMBus
On pressing in the probe the measuring value becomes smaller. ( Setting is used on probes used in
Bore gauges )
Setting for most probes with +/-1 mm, +/-2 mm measuring range. ( i.e. Tesa GT21 / GT22 )
Setting for i.e. Tesa GT61 / GT62 probes
Digital Filter off
Smallest Filter level
Highest Filter level
21