Summary of Contents for IBR IMB-im2

  • Page 1 M e t r o l o g y INSTRUCTION MANUAL IMBus IBR Measuring Bus Document No. : D2MF120 000 Edition : August 2008  Copyright : Messtechnik GmbH & Co. KG...
  • Page 2: Table Of Contents

    Instruction Manual IMBus Contents 1. Introduction 1.1 General information ......................: 1.2 Features of IMBus ....................... : 1.3 Overview over the IMBus module types ................: 2. Mounting of IMBus modules 2.1 Mechanic connection of IMBus modules ................: 2.2 Connection of IMB-Connection cable ( IMB-usb / IMB-232 ) ..........: 2.3 Connection of IMB-Connection module ( IMB-lan / IMB-wla / IMB-pb ) .......
  • Page 3: Introduction

    1. Introduction 1.1 General information The IBR Measuring Bus [ IMBus ] is a technology step in the metrology and interface technology. Powerful connection modules for all sensors and gauges as well as a maximum of flexibility by the connec- tion at current computer interfaces characterize the IMBus - Series.
  • Page 4: Overview Over The Imbus Module Types

    Instruction Manual IMBus 1.3 Overview over the IMBus module types...
  • Page 5: Mounting Of Imbus Modules

    Instruction Manual IMBus Mounting of IMBus modules 2.1 Mechanic connection of IMBus modules 1.) Push the red leavers of the first module out and turn them up. 2.) Connect the modules together. 3.) Turn red levers down to lock the modules. 4.) Connect further modules by the same way.
  • Page 6: Connection Of Imb-Connection Module ( Imb-Lan / Imb-Wla / Imb-Pb )

    Instruction Manual IMBus 2.3 Connection of IMB-Connection module ( IMB-lan / IMB-wla / IMB-pb ) 1.) Connect the IMB-Connection module ( IMB-lan, IMB-wla, IMB-pb ) to the IMBus measuring modules. ( see 2.1 ) 2.) Connect the IMB-Power supply module ( IMB-ps2, IMB-dc1, IMB-acc ) to the IMB-Connection module. ( see 2.1 ) 2.4 Mounting of bus terminator The bus terminator as well as two additional pings are part of the shipment of the IMB-Connection cables...
  • Page 7: Connection Of Hand / Foot Switches

    Instruction Manual IMBus 1.) Replace the screws at the IMBus modules by the pins included in the shipment of the IMB-Extension cables. 2.) Plug IMB-Extension cable in and screw it fix. 2.7 Connection of hand / foot switches The IMBus allows the management of several hand / foot switches. To each IMBus module a separate hand / foot switch can be connected.
  • Page 8: Power Supply

    IMB-dc1 ( DC voltage converter for input voltages 9 – 32 V DC ) 1200 mA IMB-acc ( Accumulator module for portable units ) 1200 mA 3.2 Power consumption of IMBus measuring and interface modules IMB-im1 < 20 mA IMB-im2 < 25 mA IMB-im4 < 35 mA IMB-im8 < 45 mA IMB-dm1 <...
  • Page 9: Connection Of Imbus To A Computer Or To A Plc

    Instruction Manual IMBus With the table it can be calculated, how many IMB power supply modules are needed. The IMB power supply module can be inserted at any position if necessary into the IMBus. Additional the life time of one accumulator load of the IMB-acc module can be calculated. ( Available accumulators : 1850 mAh, 4000 mAh, 5500 mAh ) On the usage of IMB-Extension cables there are brownouts on the lines.
  • Page 10 !!! Note : On the IMB-usb the Windows device driver must be installed before opening the window ( see 4.3 ) Select the PC-Connection ( i.e. COM 1, USB or IMB-lan/wlan ) and after that as IBR-instrument the IMBus or as Ser-No. the serial number of your IMB-lan / IMB-wla connection module.
  • Page 11: Start-Up Of Imb-232 Connection Cables

    Instruction Manual IMBus 3. Test of IMBus modules After pressing the Start button the test window appears : Received measurement values : - Display of transferred measuring value in the list ( which are i.e. transferred by data key of gauge ) - Display of the measuring values requested from PC ( i.e.
  • Page 12: Start-Up Of Imb-Usb Connection Cables

    Instruction Manual IMBus 2. Electrical connection : Connect the IMB-232 to IMBus as described under 2.2. Connect the free end of the connection cable to a COM-Port of the PC. Switch the IMB power supply module on. 3. Installation / Configuration : For using the IMB-232 no driver installation on the PC is necessary.
  • Page 13 The PC automatically recognizes on plugging in the IMB-usb connection cable that a driver must be installed and opens the following window : - Please insert IBR-Support-CD - Press OK b) A new window for inputting the name of the directory with the driver is opened : c) Press the Button Search..
  • Page 14: Start-Up Of Imb-Lan Connection Modules

    Instruction Manual IMBus b) Select the option Search driver : c) Select CD-ROM drive as data source and confirm the window by Next> Button. d) The driver is searched on the CD : e) By clicking on the OK Button the installation is started. 4.4 Start-up of IMB-lan connection modules 1.
  • Page 15 Instruction Manual IMBus 3. General information for connection of IMB-lan modules to a network a) On default the IMB-lan module waits for the IP address setting by a DHCP Server. If no DHCP Server is available, then the IMB-lan module automatically takes after 20 sec an IP-Address out of the address range 169.254.x.x.
  • Page 16: Start-Up Of Imb-Wla Connection Modules

    Instruction Manual IMBus - Telnet Menu „Server“ : IP-Address : IP-Address of the IMB-lan module ( 0.0.0.0 = Address setting by DHCP ) Set Gateway IP Address : IP-Address of the router, which serves as gateway for IMB-lan module This IP-Address must contain an address of the local network. Netmask : 8 = 255.0.0.0, 16 = 255.255.0.0, 24 = 255.255.255.0 Change telnet config password :...
  • Page 17 Instruction Manual IMBus Country = US Channel = 11 WEP = deactivated Data rate = 54 Mbps d) The IMB-wla modules use for the communication 3 different TCP-Ports : ( These TCP-Ports must be unlocked in the Firewall ) TCP-Port Function 0x2711 ( 10001 ) Communication with IMBus-Modules ( Measuring data, …...
  • Page 18: Start-Up Of Imb-Pb Connection Modules

    Instruction Manual IMBus - Telnet Menu „WLAN“ : Enable WLAN : By this menu point the WLAN interface of the IMB-wla module is activated / deacti- vated. This menu point MUST be set in any case to „Y“. Find Network Name : Network name of IMB-wla module Enable Ad Hoc Network Creation : Settings of Ad Hoc Network ( Network Name, Country and WLAN-...
  • Page 19 Instruction Manual IMBus 3. General information for IMB-pb modules The IMB-pb modules read continuously the actual measuring values of the connected IMB measuring modules and provide them on the Profibus. On the delivered Support-CD there is the programme IMB-PB_Setup.EXE for setting the Profibus and for firmware update of IMB-pb modules as well as the GSD-file.
  • Page 20: Configuration Of Imbus Measuring And Interface Modules 5.1 Imb-Im1/2/4/8

    Instruction Manual IMBus 5. Measuring value format of IMB-pb modules 5.1. Initialisation : After power-on the IMB-pb module waits for its initialisation by Profibus Master, this means the Profibus Master must assign the address range to IMB-pb module. Over the assigned data block size ( = address range ) the IMB-pb module knows the number of bytes to write on Profibus and calculates from this value how many IMBus measuring inputs shall be col- lected and provided on the Profibus.
  • Page 21 Instruction Manual IMBus 2. Configuration : Open Setup-Window of IMB_Test programme. Press Setup-Button behind the input ( Column Gauge / Sensor ) : Measuring direction : Standard : On pressing in the probe the measuring value becomes bigger. Inverse : On pressing in the probe the measuring value becomes smaller.
  • Page 22: Imb-Dm1/2/4

    Instruction Manual IMBus 5. Calibration with 2 masters ( = Button „Calibration“ ): The menu „Calibration” allows the calibration of the measuring inputs with 2 masters. The calling of the menu is only necessary in some special cases, because the IMB-im1/2/4/8 modules are calibrated at the manufacturer with special reference calibration equipment.
  • Page 23: Imb-Tc1/2/4

    Instruction Manual IMBus Measuring direction : Standard : On pressing in the probe the measuring value becomes bigger. Inverse : On pressing in the probe the measuring value becomes smaller. ( Setting is used on probes used in Bore gauges ) Measuring step : 0.0001 : Setting for measuring systems with 20µm period.
  • Page 24: Imb-Pm1

    Instruction Manual IMBus In addition to the IMB-sm1/2/4 modules connection cables / -adapters for the connection of the par- ticular gauges are offered. 2. Configuration : Open Setup-Window of IMB_Test programme. Press Setup-Button behind the input ( Column Gauge / Sensor ) : Selection of the gauge 5.8 IMB-pm1 1.
  • Page 25: Imb-Ci8

    Instruction Manual IMBus Electrical characteristics “Output” Operating voltage 12 V ... 32 V DC Max. output current 1 A per channel On – state resistance R 200 mΩ Current limitation 3. Wiring : 6.2 IMB-ci8 1. Short description : The IMBus input and output modules are all galvanically ( 2kV ) separated. The inputs are compatible to PLC optocoupler inputs and work in a wide voltage area.
  • Page 26: Imb-Co8

    Instruction Manual IMBus 6.3 IMB-co8 1. Short description : The IMBus input and output modules are all galvanically ( 2kV ) separated. The outputs are ESD proofed, short circuit proofed and contain a high driver power. Each in-/output contains a state LED. The connections are done by plugable terminal strips.
  • Page 27: Technical Data Of Imbus Modules

    IMBus – Module type Module width IMB-dm1, IMB-tc1, IMB-pm1, IMB-sm1 25 mm IMB-im1, IMB-ai1 37,5 mm IMB-lan, IMB-wla, IMB-im2, IMB-dm2, IMB-tc2, IMB-ae1, IMB-ai2, IMB-mi2, IMB-sm2 50 mm IMB-ps2, IMB-dc1, IMB-io4, IMB-ci8, IMB-co8 75 mm IMB-im4, IMB-dm4, IMB-tc4, IMB-ai4, IMB-mi4, IMB-sm4 87,5 mm...
  • Page 28 Instruction Manual IMBus IMB-im1/2/4/8 : ( Measuring modules for connection of inductive probes ) Supported inductive probe types All manufacturers and types Standard : Tesa Halfbridge and compatible Frequency, Input impedance, ... Corresponding to the probe specifications Resolution / measuring range 16 Bit +/- 3,0 mm on 0,1 µm +/- 30,0 mm on 1 µm...
  • Page 29 IMB-sm1/2/4 : ( Universal serial interface module ) Connection Gauges with RS232 output Universal configurable By loading gauge drivers from IBR library ( Standard : Opto-RS232 ) Operation temperature range 0 ... 45 °C Storage temperature range -20 ... 80 °C...
  • Page 30 Instruction Manual IMBus IMB-pm1 : ( Universal parallel interface module ) Connection Gauges with parallel interface ( BCD, binary, … ) Universal configurable By loading gauge drivers Adaptation of the wiring for the gauge connector By adapters par/adp1 and par/adp2 Operation temperature range 0 ...
  • Page 31 Instruction Manual IMBus IMB-acc : ( Accumulator module for portable units ) Available Accumulators 1850 mAh ( part of shipment ) 4000 mAh 5500 mAh Changing of accumulators Fast and simple Power plug from loader Can be used instead of the accumulator for power supply of the IMB-acc module Output voltage Nominal output current I...
  • Page 32: Delivered Items

    Instruction Manual IMBus Delivered items IMB-232 : ( Connection cable for IMBus to serial ports ) IMB-232 Connection cable Support-CD with drivers, IMB_Test Software and manual 4 x pins for fixing the IMB-232 and bus terminator to IMBus Bus terminator IMB-usb : ( Connection cable for IMBus to USB ports ) IMB-usb Connection cable...
  • Page 33 Instruction Manual IMBus IMB-ae1 : ( Measuring modules for connection of pneumatic gauge heads ) IMB-ae1 Measuring module Connection nibble Closing plug IMB-ai1/2/4/8 : ( Measuring modules for connection of analogue voltages and currents ) IMB-ai1/2/4/8 Measuring module IMB-mi2/4/8 : ( Interface modules for gauges with Mitutoyo Digimatic output ) IMB-mi2/4/8 Interface module IMB-sm1/2/4 :...
  • Page 34: Available Pc Software

    ). IBREXDLL ( Data collection of measuring data into MS-Excel ) With the IBREXDLL measured data can be read from all IBR interface- and measuring instruments into Excel and can be assigned freely to cells in any Excel - workbook.
  • Page 35 A 30-days Demoversion as well as further information are available under www.IBRit.com. Software packets of other companies The IBR driver is already included in many software packets and so the connection of IMBus to these software packets is possible without problems.
  • Page 36: Declaration Of Conformity

    Instruction Manual IMBus 10. Declaration of conformity Thank you very much for your confidence in purchasing this product. We herewith certify that it was manufactured and inspected in our works. We declare under our sole responsibility that this product is in conformity with technical data as specified in this instruction manual.