Zeroing the PG-6000 Gage Using the MIC TRAC (continued)
17. Turn the indicator dial on the PG-6000 to align
the needle with zero.
18. Tighten the indicator clamp.
19. Pull the retraction lever and remove the PG-
6000 from the MT-3000. Reposition the gage
on the flat face anvils to verify the zero setting.
Note: Note the position of the small revolution
counter on the indicator before removing
the gage. Place a piece of masking tape on
the backside of the indicator and record the
dial setting of the small revolution counter
to eliminate incorrect indicator readings.
20. Pull the retraction lever on the PG-6000 gage
and remove it from the MT-3000.
21. Set a frequency for verifying the zero setting of
all gages. As a minimum, the PG-6000 gage
should be zeroed on the MT-3000 once during
each shift to ensure accurate readings.