Instruction Manual Dräger P7-TDX
At the end of the stabilization period, the reduction of the test
pressure is required.
After reducing the test pressure, the measurement is started
using
.
(CONTINUE)
The pressure at the beginning of the measurement (P-Start),
the current pressure or the pressure at the end of the meas-
urement (P), the pressure difference (P-diff) and the already
elapsed time of the measurement period (t) are shown during
and at the end of the measurement.
A diagram shows the pressure curve.
The measurement can be terminated prematurely using (
).
CONTINUE
At the end of the measurement or after premature termina-
tion, a switch to the result display occurs.
A switch to the documentation menu can be made using
.