Protection Function - ITech IT6000B Series User Manual

Regenerative power system
Hide thumbs Also See for IT6000B Series:
Table of Contents

Advertisement

If you want to stop running during the battery test, you can do the following:
Press the composite keys [Shift]+[I-set] (Function) on the front panel to enter
the function menu.
At this point, the interface will prompt whether to stop the running of the Func-
tion, you can press the Left/Right key to select Stop, then the current running
will be stopped, and the Function menu item will be displayed, you can re-enter
the Function menu for editing; if Reset is selected, it means to stop the present
running and return to the main interface to wait for the next trigger operation.

5.3 Protection Function

The series instrument provides the protection functions such as overcurrent,
overpower and undervoltage protection. The corresponding protection parame-
ters can be configured in the Protect menu.
Press [Shift]+[R-set] (Protect) to enter Protect menu interface. The descrip-
tions of protect menu are listed in the table below.
Protect
Protection function menu
OCP
OPP
UVP
The trigger method of the battery test is consistent with the trigger method
of the List function, that is, the List trigger source is also effective for the
Battery function.
Overcurrent protection
Off
On
Overpower protection
Off
On
Undervoltage protection
Off
On
Copyright © Itech Electronic Co., Ltd.
Turn the OCP function off.
Turn the OCP function on.
Level
OCP limit
Delay
Delay time
Turn the OPP function off.
Turn the OPP function on.
Level
OPP limit
Delay
Delay time
Turn the UVP function off.
Turn the UVP function on.
Level
UVP limit
Delay
Delay time
Load Function