Dräger Polytron 8000 Series Technical Manual page 23

Explosion proof instrument ex d, instrument for increased safety ex e
Hide thumbs Also See for Polytron 8000 Series:
Table of Contents

Advertisement

5.2.1.3
Cable lengths for 4 to 20 mA signal connection of Polytron
Tinned copper wire, class 5/6 according to IEC 60228 / VDE 0295
Values include a safety margin of 10%.
Operating voltage: 24 V
AWG
21
0.5
2
mm
122°F
feet
906
50°C
m
276
68°F
feet
1014
20°C
m
309
Bare copper wire, class 2 according to IEC 60228 / VDE 0295
Values include a safety margin of 10%.
Operating voltage: 24 V
AWG
21
2
0.5
mm
122°F
feet
4902
50°C
m
1494
68°F
feet
1129
20°C
m
344
5.2.2
Terminal connections 4-20mA / HART
4-20mA / HART® interface
Pin
1
2
3
4
5
|
Technical Manual
Dräger Polytron
20
0.75
1361
415
1519
463
20
0.75
3002
915
1657
505
Mark
Function
PWR+
V+
PWR-
V-
PWR-
V-
To be used when the instrument is not powered by the control
unit (separate power supply), see wiring figures.
4-20mA
4 to 20 mA signal (Operation as source or sink)
PE
Protective earth
Optional PE for relay connection. Connecting PE together
with the 4-20mA analog output may cause ground loops.
®
8000 Series
18
16
1
1.5
1814
2651
553
808
2031
2963
619
903
18
16
1
1.5
2008
1483
612
452
2244
3356
684
1023
®
interface and relay
Installation
®
8200 CAT LC
13
2.5
4423
1348
4944
1507
13
2.5
1007
307
5479
1670
23

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents