Cleaning and disinfecting the foot switch
The foot switch must be cleaned and disinfected after each exam. It is important to
follow these cleaning and disinfecting instructions without skipping any steps.
WARNING:
The foot switch is not intended for use in a sterile environment.
Caution:
To avoid damaging the foot switch, do not sterilize.
Cleaning and disinfecting level
An intermediate-level of cleaning and disinfecting is required for the foot switch.
Spaulding classification
The foot switch is designed for use within the Spaulding classifications of non-critical.
Cleaning and Disinfecting - English