Cleaning; Wipe Disinfection; Wipe Disinfection With Virex Tb; Damp Heat Disinfection - Siemens Dräger Medical Carina Instructions For Use Manual

Sub-acute care ventilator
Hide thumbs Also See for Dräger Medical Carina:
Table of Contents

Advertisement

Cleaning, Disinfection and Sterilisation

Cleaning

Wipe surfaces with a clean, damp lint-free cloth.
Do not use corrosive detergents.
Aged or accumulated deposits should not be
scraped off, but should instead be softened with
warm soap solution and then removed.
Wipe until all deposits are visually removed and
allow to dry completely.

Wipe disinfection

Use preparations from the group of surface disin-
fectants for disinfection. Due to material compatibil-
ity, preparations based on these active ingredients
are suitable:
– aldehydes
– quaternary ammonium compounds.
To avoid the possibility of damage to material, do
not use any disinfectants based on:
– alkylamine-based compounds
– phenol-based compounds
– halogen-releasing compounds
– strong organic acids
– oxygen-releasing compounds.
Disinfectants often contain, besides their main
active agents, additives that can also damage
materials.
Inquiries should be referred to the supplier or
manufacturer of the disinfectant/cleaning agent.
The disinfectant in sufficient concentrations at the
correct temperature must remain in contact with the
surfaces for a specific period of time to allow pene-
tration of all the microbial cell walls and deactiva-
tion. The concentration, temperature, and exposure
times are different for each disinfectant and the dis-
infectant manufacturer's directions for use must be
followed carefully.
Chemical disinfectants should not be mixed with
each other or with detergents, since this may inac-
tivate their disinfecting properties or create noxious
fumes. Items should be dry to prevent dilution of the
disinfectant.
106

Wipe disinfection with Virex Tb

This procedure is applicable for Virex Tb. When
using Virex Tb or other disinfectants strictly follow
the disinfectant manufacturer's instructions.
Clean the surfaces according to the cleaning
instructions.
Wipe down the surfaces with a clean sponge or
cloth soaked with Virex Tb. Use sufficient disin-
fectant to saturate the surface and remain wet
for 3 minutes (for low level disinfection).
Remove Virex Tb residues with a second
sponge, cloth or paper towel.

Damp heat disinfection

93 °C/10 minutes; in the cleaning/disinfecting
machine with detergent only!
– Ventilation hoses
– Water traps and collecting jars
Autoclave sterilisation at 134 °C
– Ventilation hoses
– Water traps and collecting jars

Reprocessing of accessories

For reprocessing of the following accessories, refer
to the accompanying Instructions for Use:
– Medication nebulizer
– Respiratory gas humidifier
– Leak valve

Before reusing on patient

Assemble the device as described in the chap-
ter "Preparation" on page 31.
Check readiness for operation, see page 57.
Instructions for Use Carina SW 3.n