Setting ranges

Ventilation parameter
Inspiration pressure P
**
PEEP
[mbar]
Inspiratory O
concentration FiO
2
Trigger sensitivity Trigger
Frequency f [bpm]
Inspiration time Ti [sec]
or I:E
Ramp rise time Ramp [sec]
*
Only values can be set which are 2 mbar above the set PEEP value
**
An increase in PEEP value necessitates an increase in the Pinsp value when a min-
imum difference of 2 mbar has been reached
Instructions for Use Carina SW 3.n
*
[mbar]
insp
[%]
2
Setting range
LeakV: 5 to 40
ExpV: 5 to 50
3 to 20
21 to 100
Off/Normal/Sensitive
5 to 50
0.3 to 8
1:39 to 2:1
Auto/0.1 to 2.0
Operation
67