Safety Notices Within The Installation Instruction And The Operating Instructions For The Entire Machine; Selection And Qualification Of Personnel - Phoenix Mecano Dewert Okin DUOMAT 3 CARE EASY Installation Instructions Manual

Table of Contents

Advertisement

DUOMAT 3 CARE EASY
 You should only use spare parts which have been manufactured or approved by DewertOkin.
Only these parts will guarantee a sufficient level of safety.
Using the drive systems in medical applications
This DewertOkin product is in compliance with the safety requirements found in IEC 60601-1.
We strongly recommend that the end product (including all its components) which you are manu-
facturing for a medical application should also be in compliance with the safety requirements found
in IEC 60601-1.
You should make sure that the mechanical movement of the motor in your end product poses no
risk of injury. Conduct a risk analysis for the end product for this purpose. You should also include
safety notices in the instructions for the end product and technical safeguards in your product to
eliminate any risk.
Optional: battery-operated reset function
The battery-operated reset function is not a safety system and does not avert danger.
DewertOkin does not guarantee that the drive will function in the event of a power outage.
If the end-product manufacturer chooses to guarantee the functionality of the end product during a
power outage, then the end-product manufacturer is responsible for arranging a mechanism to en-
sure this functionality.
2.2
Safety notices within the installation instruction and the operating instruc-
tions for the entire machine
The manufacturer of the end product is only permitted to operate the DUOMAT 3 CARE EASY (by
itself an incomplete machine)
• when the end product (for which the DUOMAT 3 CARE EASY is intended) is in compliance with
all protective measures specified in the Machinery Directive 2006/42/EC, and
• when the manufacturer expressly declares the compliance of the end product.
The manufacturer of the end product must create a manual for the users of that product. The safety
notices in the end-product manual must be written based on the end product's risk assessment.
2.3

Selection and qualification of personnel

This drive should only be installed into the end product by someone who has completed training in
electronic motor assembly or has equivalent qualifications.
You should only install this drive when you are qualified to do so. Otherwise, a properly qualified
person should be found for this task.
78668(-)
CAUTION
Safety Instructions
9