Sensyflow FMT200-ECO2 THERMAL MASS FLOWMETER | OI/FMT200-ECO2-EN REV. H
Parameter for the analog output
Parameter
Analog output
Upper range value
Low flow
Filter factor
Output in the event of an error
*
not for operating mode 0-5 V, 0-10 V
** not for operating mode 0-20 mA
Description
Select the operating mode for the analog output:
0-20 mA, 4-20 mA, 0-5 V, 0-10 V
Set the flow rate for the maximum current / maximum voltage on the analog output.
Set the flow rate below which no measurement is made (measured value = 0 %)
Set the time constant for the damping of the output signal in ms.
Select the output behavior in the event of an error.
Min: Analog output 0 %
Max: Analog output 100 %
Up: Analog output > 22.5 mA*
Down: Analog output < 3.5 mA* **