Universal oct spinal stabilization

Advertisement

neon
3™
Assembly and disassembly instructions
for CS 3938-03, -04 and -05
5
neon
18
3™
Assembly and disassembly instructions
Special cleaning instructions
Cleaning the outer sleeve
Difficult to reach parts should be
pre-cleaned manually with a brush
in a water bath. In addition, the
notch at the end of the outer sleeve
should be brushed. The instrument
must then be rinsed thoroughly, to
avoid any residue from drying.
Cleaning the inner sleeve
Difficult to reach parts should be
pre-cleaned manually with a brush
in a water bath. The spring and
adjusting nut must be moved back
and forth several times. In addition,
the cannulation and the adjusting
nut should be rinsed while being
constantly turned. The instrument
must then be rinsed thoroughly, to
avoid any residue from drying.