  • Page 1 Assembly and disassembly instructions with special cleaning instructions tezo ™ titanium cage family...
  • Page 2 Notes regarding assembly and disassembly instructions This document lists all assembly and disassembly instructions available for the system tezo™. The instruments must be disassembled according to the manufacturer’s instructions to ensure the processing and storage. The product descriptions in the component overview and in the descriptive text are listed in English according to the inscription on the instruments.
  • Page 3 Contents tezo™ assembly and disassembly instructions for CS 3460-XX inserter tezo™ Assembly and disassembly instructions...
  • Page 4 tezo assembly and disassembly instructions ™ for CS 3460-XX inserter CS 3460-XX inserter, consisting of: Product number Product description CS 3460-01/-05 Inserter, length 300 mm/410 mm CS 3460-02/-06 Locking rod, length 300 mm/410 mm CS 3460-09 Adjustor CS 3460-XX inserter tezo™ Assembly and disassembly instructions...
  • Page 5 Assembly instructions Assembling the inserter Insert CS 3460-09 (Adjustor) into the cage of CS 3460-01/-05 (Inserter). Then slide CS 3460-02/-06 (Locking rod) into CS 3460-01/-05 (Inserter). CS 3460-09 (Adjustor) must be secured against falling out. tezo™ Assembly and disassembly instructions...
  • Page 6 tezo assembly and disassembly instructions ™ for CS 3460-XX inserter Disassembly instructions Disassembling the inserter By pressing against CS 3460-02/-06 (Locking rod) it can be removed from the rear from CS 3460-01/-05 (Inserter). After CS 3460-02/-06 (Locking rod) is removed, CS 3460-09 (Adjustor) can be removed from the cage of CS 3460-01/-05 (Inserter).
  • Page 7 Special cleaning instructions No special cleaning instructions. The general cleaning instructions contained in the "Processing manual implants and instruments" UH 1100 apply. tezo™ Assembly and disassembly instructions...