Assembly instructions
Assembling the rod pusher
The inner sleeve for rod pusher
(CS 3938-04) is inserted into
the outer sleeve for rod pusher
(CS 3938-05) so that the pins can
be guided into the recess. When
the first stop is reached, the inner
sleeve turns clockwise and after
releasing it, it snaps securely into
the outer sleeve.
5
For better handling, a star grip
handle (CS 3938-03) can optionally
be attached to the rod pusher.
Disassembly instructions
The rod pusher must be
disassembled for cleaning and
storage on the tray.
Removing the outer sleeve.
If the inner sleeve is turned to the
left when in a released state and
with the adjusting nut loosened,
the inner sleeve can be pulled off of
the outer sleeve.