ulrich medical Obelisc CS 2931 Series Assembly And Disassembly Instructions With Special Cleaning Instructions

Vertebral body replacement

Advertisement

Quick Links

Assembly and disassembly instructions
with special cleaning instructions
obelisc
vertebral body replacement

Advertisement

loading

Summary of Contents for ulrich medical Obelisc CS 2931 Series

  • Page 1 Assembly and disassembly instructions with special cleaning instructions obelisc ™ vertebral body replacement...
  • Page 2 Notes on assembly and disassembly instructions This document lists all of the assembly and disassembly instructions available for the obelisc system. The instruments must be disassembled according to the manufacturer's instructions in order to guarantee the processing and storage. The product descriptions in the component overview are listed in English according to the labeling on the instruments. The user must ensure that the current versions of the system-related surgical technique and the "Processing manual implants and instruments"...
  • Page 3 Contents obelisc™ Assembly and disassembly instructions for CS 2931-XX obelisc™ Assembly and disassembly instructions for CS 2931G obelisc™ Assembly and disassembly instructions for CS 2932-X obelisc ™ Assembly and disassembly instructions...
  • Page 4 obelisc™ Assembly and disassembly instructions for CS 2931-XX Components involved: Product no. Product description CS 2931-01 Holder, length 450 mm CS 2931-06 Holder, length 360 mm CS 2931-02 Expansion rod, length 450 mm CS 2931-07 Expansion rod, length 360 mm CS 2931-03 Locking rod, length 450 mm CS 2931-08...
  • Page 5 Assembly instructions Assembling the inserter CS 2931-01 or -06 is pushed over CS 2931-02 or -07. CS 2931-04 is attached to the back of CS 2931-02 or -07. At the same time, CS 2931-02 or -07 should be held from the front. Then CS 2931-03 or -08 is inserted into CS 2931-01 or -06 from the back.
  • Page 6 obelisc™ Assembly and disassembly instructions for CS 2931-XX Disassembly instructions Note: CS 2931-XX must be disassembled for cleaning and storage on the tray. Disassembling the inserter CS 2931-03 or -08 is pulled backwards out of CS 2931-01 or -06. CS 2931-04 is pulled off of CS 2931-02 or -07.
  • Page 7 Special cleaning instructions Manual cleaning All lumens and cavities must be rinsed out carefully and thoroughly. Visual inspection Finally, it must be checked whether there is any visible soiling. If this is the case, the previous steps must be repeated. Inserter maintenance For their maintenance, movable parts such as joints, hinges and...
  • Page 8 obelisc™ Assembly and disassembly instructions for CS 2931G Components involved: Product no. Product description CS 2931-10 Holder curved, length 450 mm CS 2931-11 Expansion rod flexible, length 450 mm CS 2931-12 Locking rod flexible, length 450 mm CS 2931-04 Turning knob obelisc ™...
  • Page 9 Assembly instructions CS 2931-11 is pushed into the back of CS 2931-10 curved [1]. CS 2931-04 is attached to the back of CS 2931-11 [2] and then CS 2931-12 is inserted [3]. The instrument is now fully assembled [4]. Disassembly instructions CS 2931-12 is pulled backwards out of CS 2931-10 [5].
  • Page 10 obelisc™ Assembly and disassembly instructions for CS 2931G Special cleaning instructions Important: Excessive bending of the flexible parts outside of the holder should be avoided. Manual cleaning All lumens and cavities must be rinsed out carefully and thoroughly. The flexible part of CS 2931-12 must be rinsed at the rinse holes with a disposable syringe.
  • Page 11 Instructions for machine cleaning Position CS 2931-10 at an angle by placing the handle on a tray. Position CS 2931-11 at an angle, by placing the instrument on the tray from the flexible part. General In addition, the general cleaning instructions of the Processing manual implants and instruments UH 1100 also apply.
  • Page 12 obelisc™ Assembly and disassembly instructions for CS 2932-X Components involved: Product no. Product description CS 2932-1 Handle for screwdriver, hex 3.5 mm CS 2932-2 Shaft for screwdriver, hex 3.5 mm CS 2932-3 Shaft flexible for screwdriver, hex 3.5 mm obelisc ™...
  • Page 13 Assembly instructions Assembling the screwdriver for locking screw In order to assemble CS 2932-X, CS 2932-2 or -3 is inserted into CS 2932-1 until it clicks into place. Check that the grip is secure. Disassembly instructions The screwdriver for locking screw must be disassembled for cleaning and also for storage on the tray.
  • Page 14 obelisc™ Assembly and disassembly instructions for CS 2932-X Manual cleaning All lumens and cavities must be rinsed out carefully and thoroughly. Visual inspection Finally, it must be checked whether there is any visible soiling. If this is the case, the previous steps must be repeated.
  • Page 15 Notes obelisc ™ Assembly and disassembly instructions...
  • Page 16 patented or/and pat. pend. ulrich GmbH & Co. KG l Buchbrunnenweg 12 l 89081 Ulm l Germany Telephone: +49 (0)731 9654-0 l Fax: +49 (0)731 9654-2705 spine@ulrichmedical.com l www.ulrichmedical.com...