17.4 Cleaning the Tube

(1) Remove the tube from the device and mask before cleaning.
(2) Clean the tube in warm water which contains washing liquid, and then rinse it in
clean water thoroughly.
(3) After cleaning, air-dry the tube in a cool, well-ventilated area, and avoid direct
sunlight. It takes approximately 30 minutes to completely air-dry the tube. Check
whether the tube is completely dry before re-use.
CAUTIONS!
• It is recommended to clean the tube once a week.

17.5 Replacing the Air Filter

(1) Open the air filter cap to remove the air filter.
(2) Put the new air filter in the filter area, and then place the filter cap back properly.
CAUTIONS!
• To avoid material damage, do not place the spare air filter in direct sunlight,
humid environments, or temperatures below the freezing point. The air filter should
be replaced every 2 weeks (It may be replaced more frequently based on actual
sanitary conditions).
• Operating the device with a dirty air filter may stop it from working properly and
may cause damage to the device.
• Replacing the air filter and filter cap, device must be unplugged.

18. Traveling with the Device

CAUTIONS!
• Empty the water chamber before packing the device for your trip; in order to
prevent any remaining water from entering the device.
• Using the device at an incorrect elevation setting could result in airflow pressures
higher than the prescribed setting. Always verify the elevation setting when
traveling or relocating.
• If the device is used when the atmospheric pressure is out of the stated range (See
Section 5), the accuracy of the leakage alert will be affected.
(1) Use the 3B medical, Inc. carrying case to carry the device and accessories along
with you. Do not put them in your checked baggage.
(2) This device operates on power supplies of 100 ~ 240 V and 50 / 60 Hz, and is
suitable for use in any country in the world. No special adjustment is necessary, but
you will need to find out the types of the power sockets in your destination. Bring, if
necessary, a power socket adaptor which can be purchased in electronics stores.
(3) Remember to bring a spare air filter and the emergency documents (filled and
signed by your physician) about this device. If you plan to travel by air, remember to
bring the multi-language emergency documents about respiratory therapy, in case