Cirrus 5 User Manual contained in this manual before operating your oxygen concentrator! Contents Introduction…………………………………………………………3 Symbols…..…………………………………………………4 Safety Notice……....………………………………………5 Product Introduction…........……..…………7 Operation Conditions and Environment…………………………7 Indications for Use…………..……………………………7 Technical Parameters………..………..………………………8 Structures and Functions...…………..………………………9 …………..........…10 Operation Instructions 10 Alarms and Safety Devices..………..…………………13 Maintenance…...………………………..……………14...
Cirrus 5 User Manual 1 Introduction Thank you for purchasing the Cirrus 5 Oxygen Concentrator. This user manual contains function, operation steps, basic trouble shooting and important safety information. To ensure your safety and proper operation of this oxygen concentrator,...
Cirrus 5 User Manual 2 Symbols The following table is a list of symbols and definitions regarding the Oxygen Concentrator. Symbol Description Symbol Description Warning – Describes a Indicating its conformity with the hazard or unsafe practice Medical Devices Directive that can result in severe 93/42/EEC.
Cirrus 5 User Manual Safety Notice : This unit is not a life-support device, and in certain circumstances oxygen therapy hazardous, it is suggested that if any patient who needs oxygen treatment, please follow your physician’s advice to choose the right flow and period for oxygen before using the oxygen concentrator.
Page 6
Cirrus 5 User Manual : Do not place the oxygen concentrator in surroundings where its airflow is obstructed. : Do not place items on top of the concentrator. : Always place the concentrator on a hard surface. Never place the concentrator on a surface such as bed or couch, where the concentrator may tip or fall.
A high concentration of supplemental oxygen is supplied and a nasal cannula is used to channel oxygen from the concentrator to the patient. The Cirrus 5 can be used in a home, institution, vehicle and various mobile environments.
Structures and Functions Figure 1 Figure 2 1. Indicating Lamp Total 6 indicating lamps and their indication for the Cirrus 5 oxygen concentrator are as follows: a. P.O.: power switch (green lamp) b. P.F.: power failure (red lamp) c. L.P.: low pressure (yellow lamp) d.
Cirrus 5 User Manual 6. Intake air filter 7. Storage cabinet For storage of oxygen cannula 8. LCD display (Liquid crystal display) a. Displays operating status of the oxygen concentrator, refer to 7. on page 9. b. When starting the oxygen concentrator, the LCD screen will be backlit. It will return to...
Page 11
Cirrus 5 User Manual Figure 3 Figure 4 9.2. Connect the nasal oxygen cannula to the humidifier outlet connector or to the concentrator outlet if a humidifier is not is use. The nasal oxygen cannula and Oxygen tubing should be limited to 57 feet in length, in order to ensure that the oxygen flow rate remains within specification values.
Page 12
Cirrus 5 User Manual Flow value: Cirrus 5:10 position flow value from 0.5 ~ 5L/min on flowmeter as shown in figure. The maximum recommended flow: 5L/min. In compliance with the ISO8359 standard, the flow supplied is equal to the flow set on the flowmeter, accurate to within ...
Cirrus 5 User Manual Alarms and Safety Devices 10.1 Alarms a. Power failure alarm: In case of a loss of main power or when the power cord is not plugged into the wall outlet, an audible alarm is activated with red indicator on (8.1b on page 10).
Cirrus 5 User Manual Maintenance Disconnect the power cord from the electrical outlet before you clean the cabinet (Fig12). Do not operate the concentrator without the filters installed, or while filters are wet. These actions could permanently damage the concentrator.
Page 15
Cirrus 5 User Manual 11.3. Clean secondary filter First, remove the silicone tube connected with oxygen output, take down the humidification bottle, rotate the air filter with a counterclockwise direction as Figure 14, remove the filter cloth, clean with mild soap, rinse with water, wipe off the water. Air dry, then install back to the air filter.
Cirrus 5 User Manual The other fuse is located at the intake of internal power line. Follow above instructions to replace. Figure 15 Troubleshooting If your concentrator fails to operate properly, please refer to the troubleshooting chart on the following pages for probable causes and solutions.
Accessories, transducers and cables may be specified either generically or specifically. NOTE: Transducers and cables sold by the manufacturer of the Cirrus 5 as replacement parts for internal components need not be listed. The use of accessories, transducers and cables other than those specified, with the...
Page 18
Cirrus 5 User Manual Guidance and manufacturer’s declaration – electromagnetic emissions The Cirrus 5 is intended for use in the electromagnetic environment specified below. The customer or the user of the Cirrus 5 should assure that it is used in such an environment.
Page 19
Guidance and manufacturer’s declaration – electromagnetic immunity The Cirrus 5 is intended for use in the electromagnetic environment specified below. The customer or the user of the Cirrus 5 should assure that it is used in such an electromagnetic environment.
Page 20
If the measured field strength in the location in which the Cirrus 5 is used exceeds the applicable RF compliance level above, the Cirrus 5 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Cirrus 5.
Cirrus 5 User Manual Accessories Your concentrator includes the following components: • Intake air filter (two pieces, part number:GL-01) • Secondary filter (one piece, part number:GL-02) The concentrator comes with two air filters and one secondary filter already installed. : Please use the parts mentioned in this chapter. The use of other parts can degrade minimum safety and performance.