Luna II QX CPAP / Luna II QX Auto CPAP User Manual V1.1
Luna II QX CPAP / Luna II QX Auto CPAP User Manual V1.1
iCode
iCode QR,
iCode QR +
0 ~ 50000
Use Time
About

16. Alert

Alert Message
Power Failure!!!
Device Fault!!!
Leak!!
Low Input
Voltage!!
Humidifier
Failure!!
Please Change
Filter!
SD Card Full!
Reinsert SD card!
"Nasal," but the patient can choose other
suitable masks as well. When selecting masks
other than the above three types of 3B
Medical, Inc. masks, the patient can identify
the masks as other.
iCode
iCode,
compliance data during a recent time period.
The iCode mode displays data in sequences of
numbers, and the iCode QR / iCode QR +
mode displays data in two-dimensional codes
Use Time displays how long has the device
been used by the patient
h
Displays related information of the device
——
(Model, SN, Version, ID). This is read-only and
cannot be edited
An audible alert will sound in 6 s if the device is accidentally
disconnected from power when it is delivering air.
Note:
(1) The alert will not sound if power failure occurs when the
device is in standby state.
(2) No alert message on the screen during a power failure
An audible alert will sound if no airflow comes out of the
machine; the screen will display "Device Fault!!!"
When the airflow is on, an audible alert will sound in 40 s if the
air leak rate is excessive; the screen will display "Leak!!"
If the voltage supplied by power adaptor is lower than 22V,
an audible alert will sound and the screen will display "Low
Input Voltage!!"
When humidifier is applied, an audible alert will sound when
the humidifier fails to work; the screen will display "Humidifier
Failure!!"
When the Filter Alert feature is enabled, an audible alert will
sound if the preset replacement time reaches but without
replacing the air filter; the screen will display "Please Change
Filter!"
The screen will display "SD Card Full!" if the SD card has
reached its maximum capacity
The screen will display "Reinsert SD card!" if the SD card fails
to work
26