Chapter 4: Diagnostics and Troubleshooting

Error Number 86

This error condition indicates the BBU RAM Battery is Low. Report
this error to your Service Department.

Error Numbers 92-100, 99, 100, 111-114, 116

These errors pertain to Dud Container Inspection. Report any of
these errors to your Service Department.

Error Number 117

This error condition indicates there is a Host or MODBUS Serial
Interface Framing Error or Reply Timeout (timeout must be a
minimum of 15 seconds.) If this error persists, report this error to
your Service Department.

Error Number 118

This error indicates a host or Modbus serial RX overrun error. The
baud rate may be set incorrectly. If this error persists, report this
error to your Service Department.

Error Number 119

This error indicates a host or Modbus serial parity error occurred.
If this error persists, report this error to your Service Department.

Error Number 120

This error indicates there is a RX Receive Rate error. Too many
characters have been received in one second. If this error persists,
report this error to your Service Department.

Error Number 121

This error indicates the CTS line (Clear To Send) is changing too
rapidly for normal communication and therefore appears busy. If
this error persists, report this error to your Service Department.

Error Number 123

The X-ray Status Function is turned off. Turn the status back on to
activate x-ray generator.