System Functions Group - Filtec FT-70 Operation Manual

Fill level inspectors
Table of Contents

Advertisement

FT-70/75 Fill Level Inspectors Operator's Guide
3-14

System Functions Group

Current Container Type
The container type that is currently being inspected. This function
allows you to select a stored configuration for a particular
container type. If your production line is set up for more than one
container type, this function allows you to switch to different
container production without having to reconfigure the software
function parameters.
This function is only available if the unit is configured for multiple
container types. For more information see Chapter 2, "Container
Changeover".
To select the Current Container Type:
1. Press Key 5 until CURRENT CONTAINER TYPE appears.
2. Press the Arrow Key until the desired container type appears.
Last Reject Detected
The type of failure that caused the most recent reject container. The
failure classification can be any of the inspection capabilities
installed and active on your system.
Beacon Status
Determines the operational status of the warning beacon. There are
three status choices:
• On: the beacon will flash once for each rejected container and
continuously when a diagnostic error occurs. The beacon will
stop flashing when the diagnostic error is cleared.
• Off: The beacon will flash once for each rejected container.
• High Priority Only: The beacon will flash once for each
rejected container and continuously when a high priority
diagnostic error occurs. The beacon will stop flashing when the
diagnostic error is cleared.
Industrial Dynamics Company, Ltd. Document 40274-0600