Procedure - Keysight N5221A Security Features And Document Of Volatility

Pna network analyzers
Hide thumbs Also See for N5221A:
Table of Contents

Advertisement

Memory Sanitization and Removal Procedures
Instrument Sanitization Procedure

Procedure

Steps below that appear on a yellow background take place in the controlled
area. All other steps take place in a non-controlled area.
Step
1
2
3
4
5
6
7
8
9
10
11
12
13
Security Features & Document of Volatility
Task
Receive new instrument from factory.
Power up the instrument and verify the Instrument Software Revision.
For instructions on how to verify the currently installed Instrument Software Revision,
see
"Is your product software up-to-date?" on page
Write down the software revision.
If the instrument's software is not the latest version, update the instrument software to
the latest version available. For instructions, see
up-to-date?" on page 3
Write down the software revision.
Replace the original instrument disk drive (Disk Drive #1) with the spare disk drive
(Disk Drive #2) that was obtained previously.
For details of how to remove the Disk Drive, see
on page
25.
Turn on the instrument with the spare Disk Drive #2 installed.
Verify that the instrument software on the spare Disk Drive #2 has the same revision as
that on the original Disk Drive #1.
If the Instrument Software Revision on the spare Disk Drive #2 is not the same as that
on the original Disk Drive #1, update the instrument software on this drive to match the
software revision that you documented in step 3.
For instructions, see
"Is your product software up-to-date?" on page 3
Write the instrument model number, serial number, and software revision on the
original Disk Drive #1, then reinstall it in the instrument.
Write the instrument model number, serial number, and software revision on the spare
Disk Drive #2, place it back in the static safe bag that it was shipped in, then store it in
a safe place until it is needed again.
The instrument is now ready for use in a secure environment.
Physically deploy instrument into controlled area.
Use instrument inside controlled area until cal or repair is needed.
Make a note of the current software version, because the instrument's software may
have been updated inside the secure area. To determine this, go to the main menu and
select Help > About PNA.... Write down the application code and disk drive revision
and attach this note the instrument.
3.
"Is your product software
"Disk Drive Removal Procedure"