® Service Manual for RESmart Auto CPAP V1.1 Table of Contents 1. Introduction................................1 1.1 Intended Use ..............................1 I.2 Implement............................... 1 2. Structure................................. 1 2.1 Outer Illumination ............................1 2.2 Inner Construction ............................2 2.3 Spare Part............................... 2 3. Software ................................. 2 3.1 Instruction for Upgrading Device Software....................
Page 3
If applying this manual on other purpose, the validation should be confirmed by BMC Medical Co. Other wise, BMC Medical Co. has no responsibility for the result. All information in this manual is protected by law and regulation. All copyrights are belongs to BMC Medical Co.
® This service manual is used to help service people maintenance RESmart Auto CPAP system of BMC Medical Co. more efficiently. Instructions in this manual may help device work best. Engineer can find needed instruction from this manual quickly. IMPORTANT! Read and understand instruction in this manual before operating.
Auto CPAP device, then connect PC and the device via data cable. ® 2. Run the software “BMC RESmart Tools for Upgrade” from PC and display as below: a) Select the upgrade file which has same suffix letter as the machine version. (For example V ersion 1.88-EN should choose file name 1.98-EN)
® Service Manual for RESmart Auto CPAP V1.1 ® Power on the RESmart Auto CPAP device. If the device is power on from beginning, please power off the device and turn on again after 5 seconds more. 4. After the device is turned on, the following will display: ®...
® Service Manual for RESmart Auto CPAP V1.1 ‘YYYY/MM/DD’=’2098/08/18’, press ‘Pressure Start/Stop Button’ when Hour Setting appears to start the Pressure Calibration. The device output pressure is based on 20 hPa (cmH O), and a number will appear on the screen, when the output pressure is stable, if the value on the pressure meter is not 20 hPa (cmH O), press ‘+/-’...
® Service Manual for RESmart Auto CPAP V1.1 Appendices Appendix A: Device Configuration 1. Main device Fig. A - Main board (backside) Fig. B – Outlet (leftside) Hole #1, 2 and 3 in Fig. B, connect to sensor YL and LL via silicon rubber canal. Specified in Table 1: Fig.
Page 10
® Service Manual for RESmart Auto CPAP V1.1 Fig. C – Power PCBA (front side) Sockets figuration Fig. D – Wire connecting on power PCBA Fig. E – Power supply connecting 7/16...
® Service Manual for RESmart Auto CPAP V1.1 Fig. F – Power supply to humidifier Connecting relationship: Part Fig. Back of device Fig. C: CN1 Fig. C: CN2 Fig. A: P24 Fig. F: Left Fig. F: Right Blower 8-pin plug Fig.
Page 12
® Service Manual for RESmart Auto CPAP V1.1 Fig. H – Power supply to humidifier Right Left Orange Black Fig. I – Heater PCBA Connecting relationship: Part Fig. Fig. G: P2 Fig. G: P1 Fig. H: Left Fig. H: Right Table 3 9/16...
® Service Manual for RESmart Auto CPAP V1.1 Components Illumination 1. Device Inside Assembly Figure Insert humidifier locker on the shield bottom. Insert humidifier power supply reed on the shield bottom as per Fig. E and F. Assembly the humidifier locker and button spring.
Page 14
® Service Manual for RESmart Auto CPAP V1.1 Connect blower to connector and fix by clip. Put the integrated one in the right place. Covered by foam cover and clean up wires. Assembly fixing clip on the main board PCBA. Plug the 8-pin plug of blower in the socket PM plug #2 as in Fig.
® Service Manual for RESmart Auto CPAP V1.1 2. Device Outside Assembly Figure Put handle in the socket on shield cover. Assembly handle fixing to handle and fix by screw (4×3mm*8mm). Assembly the infrared window. Assembly and fix the panel. Put on the key button.
® Service Manual for RESmart Auto CPAP V1.1 Put spring on each of three poles on heater plane. Assembly the scaleboard on the heater plane and fix by screw (3mm*8mm). Scaleboard As per Table 3, assembly heater PCBA and wires. Put the humidifier inner connector on the shield.
Page 17
® Service Manual for RESmart Auto CPAP V1.1 Humidifier Clip Humidifier Locker Humidifier Fixing Filter Cover Shield Back Outlet PCBA Fixing Infrared Window Panel Humidifier Outer Shield Humidifier Inner Shield Light Window 14/16...
® Service Manual for RESmart Auto CPAP V1.1 Pole Platelet Humidifier Power Plug Humidifier Plug (left-up) Humidifier Plug (left-down) Humidifier Plug (right-up) Humidifier Plug (right-down) Device Pad Blower Connector Humidifier Inner Connector Key Button Humidifier Scaleboard Heater Plane Spring Button Spring Spare Part List Part Name Unit...
Page 19
® Service Manual for RESmart Auto CPAP V1.1 Heater Plane 230006 Assembly Humidifier Power Supply PCBA Wire 120002 Wire Main Device Main Board PCBA Wire 120003 Wire Main Device Silicon Rubber Canal 290001 Assembly Main Device Foam (bottom) 240001 Foam Main Device Foam (cover) 240002...